MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION COMPACT
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 07/06/2017
Event Type  malfunction  
Manufacturer Narrative
Additional information: system serial number is now provided. A full system check-out was completed and all tests passed. Full system functionality was confirmed and the system was returned to service.
 
Manufacturer Narrative
The computer for the navigation system was received by the manufacturer for evaluation. When loading large exams, it was found that the computer would lag. When loading smaller exams, the computer functioned as designed. The hardware investigation found that the reported event was related to a hardware issue. This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Manufacturer Narrative
Additional information: device manufacturer date provided. The software investigation found that a probable cause was unable to be determined without further information since the on-going investigation proved to be inconclusive based on the information provided.
 
Manufacturer Narrative
Correction: the software investigation found that the reported event was unrelated to a software issue. The software functioned as designed.
 
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. No parts have been received by the manufacturer for evaluation. Part not returned for analysis.
 
Event Description
A medtronic representative reported that, while outside of a procedure, the navigation system became unresponsive when loading patient data into the application software. It was reported that the application software was reinstalled without resolution. There was no patient present when this issue was identified. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMEDTRONIC FUSION COMPACT NAVIGATION SYSTEM
Type of DeviceEAR NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key6763553
MDR Text Key102877301
Report Number1723170-2017-03174
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION COMPACT
Device Catalogue Number9735602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-