GUIDANT PUERTO RICO BV INGENIO; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number V273
Device Problem Over-Sensing (1438)
Patient Problem No Information (3190)
Event Date 04/01/2017
Event Type  malfunction  
Manufacturer Narrative
With the exception of model numbers, all other pertinent patient and product information was de-identified.As no further information concerning this report is expected, our investigation is complete.
 
Event Description
In the interest of continuous improvement, a study was undertaken by boston scientific¿s research and development to evaluate potential device enhancements to detect a phenomenon known as ¿mv/rrt noise¿.The work from this research study may be subsequently used to create a new feature in the device that detects and resolves this mv/rrt noise issue.A large data set of episode electrograms from boston scientific¿s latitude remote monitoring system will be acquired.Mv/rrt noise in the data set will be characterized.Once characterized, proposed device enhancement features will be tested.Based on review of stored data by the adjudication committee involving 5 model#v273 devices, the following clinical observation was identified in some of the episodes: oversensing (greater than 2 seconds of asystole).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENIO
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT PUERTO RICO BV
Manufacturer (Section G)
GUIDANT PUERTO RICO BV
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key6768291
MDR Text Key81855023
Report Number2124215-2017-14045
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial
Report Date 04/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberV273
Other Device ID NumberINTUA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-