The replaced oxygen gas module was returned together with a complete set of device logs no damages or deviations were found during a visual inspection of the returned gas module.The gas module was tested i a reference machine but the reported issue with the oxygen concentration not being measured according to settings has not been reproduced during this investigation.The tested oxygen gas module worked as intended during the 4 days long test period.An evaluation of the received device logs show that no technical alarms were generated during the event.The logs show that the fio2 level was lower than set at some points but the measured fio2 value and set o2 concentration are not expected to match at all times in an anesthesia machine in manual gas control mode (mgc) since when there is a certain amount of gas that is being re-breathed via the 1,2 liter large volume reflector (the tidal volume was around 80ml) it takes time for the oxygen concentration to stabilize and that is probably why the customer reacts to the ¿lower fio2¿ than set.Based on the investigation of the returned o2 gas module and the evaluation of the device logs, our conclusion into this matter is that the experienced "lower fio2 than set" was not caused by a device malfunction.
|