Model Number 174006 |
Device Problems
Failure to Advance (2524); Mechanical Jam (2983)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/29/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a rectal prolapse surgery, the device was jammed in the shaft and did not come out.Same event occurred in two devices in same case.The procedure was completed with another device.The status of the patient is no problem.
|
|
Manufacturer Narrative
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted the unit had disrupted timing.The handle was unable to be actuated.The shaft was disassembled from the handle and the handle was able to cycle properly.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the reported condition is caused by an instrument that has been exposed to excessive force while applying helixes to a surface.If a helix is fired over improper surfaces it can provoke the exertion of excessive force to the handle causing the unit to disrupt the timing and to a possible jam.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|