• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22495
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/21/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4)2017, that on (b)(6)2017, the receiver was overheating.No additional event or patient information is available.The device has been received for evaluation.  a follow-up report will be submitted once the evaluation is complete. .
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint receiver device was returned for evaluation.An external visual inspection was performed and no defects were found.The receiver was charged and rebooted.The receiver log was downloaded and evaluated with no related errors observed.Global functional testing was performed and passed.The receiver case was opened for an internal visual inspection and it passed.The reported event of an overheating or shocking device was not confirmed.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6785477
MDR Text Key83148069
Report Number3004753838-2017-64104
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10386270000153
UDI-Public(01)10386270000153(241)STK-DR-001(10)5223529
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495
Device Catalogue NumberSTK-DR-001
Device Lot Number5223529
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
-
-