U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. ARROW PI CVC KIT: 3-L 7 FR X 20 CM AGB; CATHETER,INTRAVASCULAR,THERAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. ARROW PI CVC KIT: 3-L 7 FR X 20 CM AGB; CATHETER,INTRAVASCULAR,THERAP Back to Search Results
Catalog Number CDC-45703-XP1A
Device Problems Bent (1059); Difficult to Insert (1316); Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer alleges that the doctor could not insert the spring wire in the course of action the wire bent.The entire wire was removed from the patient without an event.
 
Manufacturer Narrative
(b)(4).The customer returned a single guide wire for evaluation.The guide wire was returned within the advancer tube and showed evidence of use.The guide wire was observed to have a gradual bend/kink towards the distal end of the body.The distal j-bend was slightly deformed but intact.Microscopic examination confirmed the kink in the guide wire body.Both welds were present and were observed to be full and spherical.The guide wire was kinked/bent from 3.1-4.2cm from the distal tip.The overall length and outer diameter (od) of the guide wire were found to be within specification.The guide wire was advanced through a lab inventory ars and a lab inventory 18ga introducer needle to functionally test the guide wire.The guide wire passed through both components with minimal resistance.A manual tug test confirmed both the distal and proximal welds are intact.A dhr review was performed on the guide wire and insertion components and no relevant manufacturing issues were identified.The instructions-for-use (ifu) provided with the kit describes suggested techniques to minimize the likelihood of guide wire damage during use.The instructions caution that withdrawing the guide wire against the needle bevel or use of excessive force during removal could damage or break the wire.The report that the guide wire kinked during use was confirmed through examination of the returned sample.The guide wire had a bend/kink between 3.1-4.2cm from the distal tip.The returned guide wire met all relevant dimensional requirements.A dhr review was performed and did not identify any manufacturing related issues.Based on the condition of the guide wire and the report that the damage was observed during use, it was determined that operational context caused or contributed to this event.
 
Event Description
The customer alleges that the doctor could not insert the spring wire in the course of action the wire bent.The entire wire was removed from the patient without an event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW PI CVC KIT: 3-L 7 FR X 20 CM AGB
Type of Device
CATHETER,INTRAVASCULAR,THERAP
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key6795390
MDR Text Key82782158
Report Number1036844-2017-00305
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Catalogue NumberCDC-45703-XP1A
Device Lot Number23F17D0388
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received09/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-