Model Number MERGE PACS V7,1,2
Device Problem Computer Software Problem (1112)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/20/2017
Event Type  malfunction  
Manufacturer Narrative
After further investigation, development personnel determined this was a defect in the code that skips corrupted images.The code was changed in merge pacs version 7.1.2.1.The condition of skipping corrupted images was removed and a load error is displayed to a user while scrolling through a study with corrupted images.The user may be able to re-send the study images from the modality without any corruption which would then allow for scanning through images without issue.
 
Event Description
Merge pacs is a picture archiving communication system that is intended to create and display two-dimensional and three dimensional images of anatomy from a series of digitally acquired images.Pacs is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities.On (b)(6)2017, a customer identified an issue with image load errors while testing merge pacs version 7.2.1 in their test environment.It was determined that images that failed to load properly were corrupted.The corrupted images were skipped during manual scrolling through a study.There is no warning or error that indicates images have been skipped.There is a potential that a skipped image may contain data that is required to determine diagnosis and/or treatment decisions.Without necessary information, there is a potential for a delay in diagnosis &/or treatment or an inaccurate diagnosis &/or treatment of a patient.At this time, there is no evidence of patient harm or delay due to this issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE PACS
Type of Device
PICTURE ARCHIVING COMMUNICATION SYSTEM PACS
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge dr
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge dr.
hartland WI 53029
Manufacturer Contact
meg mucha
900 walnut ridge dr.
hartland, WI 53029
2629123514
MDR Report Key6796510
MDR Text Key83014521
Report Number2183926-2017-00168
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082144
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE PACS V7,1,2
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-