Merge pacs is a picture archiving communication system that is intended to create and display two-dimensional and three dimensional images of anatomy from a series of digitally acquired images.Pacs is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities.On (b)(6)2017, a customer identified an issue with image load errors while testing merge pacs version 7.2.1 in their test environment.It was determined that images that failed to load properly were corrupted.The corrupted images were skipped during manual scrolling through a study.There is no warning or error that indicates images have been skipped.There is a potential that a skipped image may contain data that is required to determine diagnosis and/or treatment decisions.Without necessary information, there is a potential for a delay in diagnosis &/or treatment or an inaccurate diagnosis &/or treatment of a patient.At this time, there is no evidence of patient harm or delay due to this issue.(b)(4).
|