The ecm has not been returned to isi for evaluation.Therefore, the root cause of the customer reported failure mode cannot be determined.A follow-up mdr will be submitted if the instrument is returned (post engineering evaluation) or if additional information is received.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if this malfunction were to recur it could cause or contribute to an adverse event.
|
It was reported that during a da vinci-assisted surgical procedure, the customer experienced a repeated recoverable fault 23017 on the endoscopic camera manipulator (ecm).An intutive surgical, inc.(isi) technical field specialist (tfs) was present during the procedure when the error occurred.However, the surgeon made the decision to convert and complete the procedure as a traditional open surgical procedure as the patient was bleeding and did not have time for troubleshooting.There was no report of patient harm, adverse outcome or injury.After the procedure was completed, the tfs found the error in the log files pointing to ecm 1 axis 4.The tfs replaced the ecm to resolve the error.The ecm is the camera arm located on the patient side cart (psc) which provides the sterile interface for the 3d endoscope.
|