MAUDE Adverse Event Report: WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-1-JUG-TULIP
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/17/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation is still in progress.
 
Event Description
Description of event according to initial reporter: deployment difficulty occurred during the procedure.Patient outcome: unknown.
 
Manufacturer Narrative
Exemption number e2016032.(b)(4).Name and address for importer site: (b)(4).Summary of investigational findings: investigation is based on event description only.No product returned and without the actual complaint device it would be inappropriate to speculate at what may or may not have led to the difficulties in deploying/releasing the filter.However, it is known from the literature that excessive back tension could result in deployment difficulties/failure when pressing the release button.The ifu addresses the issue through the statement "while keeping slight back tension on the introducer, push the release button completely to ensure proper release of the filter" and the warning "excessive tension during deployment may prevent the filter from releasing when the release mechanism is activated".No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that it did not perform as intended.Cook medical will continue to monitor for similar events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer Contact
thomas hessner kirk
sandet 6
bjaeverskov DK-46-32
DA   DK-4632
56868686
MDR Report Key6799845
MDR Text Key83208940
Report Number3002808486-2017-01631
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-1-JUG-TULIP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date07/24/2017
Date Manufacturer Received11/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-