• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T GEN 2 BIBAG WITHOUT CDX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T GEN 2 BIBAG WITHOUT CDX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190895
Device Problem Pumping Stopped (1503)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/26/2017
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation. Additionally, no on-site evaluation of the unit was performed by a fresenius regional equipment specialist (res). The biomedical technician (biomed) replaced the blood pump module which resolved the issue. The unit was returned to service at the user facility without a recurrence of the event as reported. A records review was performed on the reported serial number. An investigation of the device manufacturing records was conducted by the manufacturer. There were no non-conformances or any associated rework during the manufacturing process which could be associated with the reported event. In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements. The investigation into the cause of the reported problem was not able to be confirmed. A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A user facility biomedical technician (biomed) reported that the patient experienced blood loss while undergoing hemodialysis (hd) treatment on a 2008t hd machine. The biomed reported that the blood pump failed during treatment and the total estimated blood loss (ebl) was noted as being approximately 200 milliliters (ml). Follow-up with the nurse revealed that the blood pump stopped working approximately two and a half to three hours into the treatment. The machine alarmed for high pressure shortly before the blood pump stopped but the machine did not alarm when the pump failed. The staff was unable to return the patient¿s blood via manual pump. The patient¿s blood in the circuit was discarded. There was no reported defect or damage observed with the dialyzer or bloodlines and no report of blood clotting. The hd treatment was terminated early and was not resumed. No patient adverse effects were experienced and no medical intervention was required as a result of this event. The patient has since returned to the facility for regularly scheduled hd treatments with no further issues. The machine was removed from service following the event for further evaluation by the on-site biomed. The biomed replaced the blood pump module which resolved the issue. The unit was returned to service at the user facility without a recurrence of the event as reported. No parts are available to be returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008T GEN 2 BIBAG WITHOUT CDX
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6800272
MDR Text Key249094269
Report Number2937457-2017-00740
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/06/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number190895
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device AgeMO
Event Location No Information
Date Manufacturer Received09/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-