• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-V2
Device Problems No Display/Image (1183); No Device Output (1435)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/20/2017
Event Type  malfunction  
Manufacturer Narrative
This device referenced in this report has not been returned to olympus medical systems corp.For evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available at a later time, this report will be supplemented.
 
Event Description
Olympus was informed that the subject device did not display image when it was used for an unspecified therapeutic procedure.It was also reported that a pin, one of the component of the venting connector, was missing and there was rust at the venting connector.There was no patient injury reported.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.But was returned to olympus (b)(4).It was confirmed during performed investigation that the malfunction of the image, the connecting tube was heavy kinked, and the venting connector¿s pin was broken.The manufacturing record of the subject device was reviewed with no irregularity.
 
Manufacturer Narrative
This supplemental report is submitting to correct "device product code".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
katsuaki morita
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key6802975
MDR Text Key83222462
Report Number8010047-2017-01212
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCYF-V2
Other Device ID Number04042761066778
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/13/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-