• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT PUERTO RICO OPERATIONS CO SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT PUERTO RICO OPERATIONS CO SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problem Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/16/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer's representative (rep) regarding a patient who was receiving 25mg/ml dilaudid at 1.379mg/day via an implantable infusion pump for malignant pain.It was reported that a motor stall was seen at the initial interrogation of the pump and that the patient recently had a magnetic resonance imaging (mri) scan.The mri was not done due to a problem with the device or therapy.The motor stall had not recovered and it had been 2.5 hours since the mri at the time of the report.The rep stated they will periodically interrogate the logs to see if the stall had recovered.No further complications were anticipated/reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6804228
MDR Text Key83045943
Report Number3004209178-2017-17715
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00613994779229
UDI-Public00613994779229
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Notification
Type of Report Initial
Report Date 08/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2014
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Date Manufacturer Received08/16/2017
Date Device Manufactured07/21/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0591-2009
Patient Sequence Number1
Patient Age31 YR
-
-