• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P 3DKNEE; INSERT, 3D EX SZ 8LT 9MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P 3DKNEE; INSERT, 3D EX SZ 8LT 9MM Back to Search Results
Model Number 391-09-708
Device Problem Failure to Align (2522)
Patient Problem No Code Available (3191)
Event Date 07/20/2017
Event Type  Injury  
Event Description
Revision surgery - due to the patient experiencing tibia mal-alignment.
 
Manufacturer Narrative
The reason for this revision surgery was the patient had mal-alignment of the tibia insert.The in-vivo length of patient service for the implant was 2.1 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed to be non-product related.The root cause for this event was the patient had mal-alignment of the tibia insert.The root cause for the tibia insert mal-alignment was not reported.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions, root cause, relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DKNEE
Type of Device
INSERT, 3D EX SZ 8LT 9MM
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd.
austin, TX 78758-5445
MDR Report Key6805853
MDR Text Key83092476
Report Number1644408-2017-00691
Device Sequence Number1
Product Code OIY
UDI-Device Identifier00888912118989
UDI-Public(01)00888912118989
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091956
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model Number391-09-708
Device Catalogue Number391-09-708
Device Lot Number59605229
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
-
-