• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT6
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
Product evaluation: the product was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Additional information has been requested.The manufacturer internal reference number is: (b)(4).
 
Event Description
A facility representative reported an intraocular lens (iol) that was exchanged due to "wrong size moved" following a procedure.Additional information has been requested.
 
Manufacturer Narrative
The lens was returned in a specimen cup.Solution is on the lens.Both haptics are slightly bent in the distal areas.The optic is torn/split/cracked and cut, typical of insertion and removal.The cut optic portions have multiple scratches.Power and resolution testing could not be conducted due to the extensive optic damage.Power and resolution testing could not be conducted due to the extensive optic damage.Product history records were reviewed and the documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint of ¿exchanged due to wrong size "moved" ¿.Lens damage was observed typical of insertion and removal.File information indicates the lens was explanted and exchanged.The replacement lens model and diopter are unknown.The position of the lens while in the eye cannot be determined.A lens bench evaluation could not be conducted due to the lens damage.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6807801
MDR Text Key83152397
Report Number1119421-2017-01062
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberSN6AT6
Device Catalogue NumberSN6AT6U145
Device Lot Number12504418
Other Device ID Number00380652274753
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received10/23/2017
Date Device Manufactured11/15/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-