The external visual inspection revealed that the silastic overmold was cut at the electrode lead.This is believed to have occurred during revision surgery.The photographic imaging inspection confirmed severed electrode wires.This is believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed all of the electrical tests performed.The device failed the residual gas analysis test.The reported complaint of patient decrease in performance could not be verified during this analysis, which was limited in some respects due to the condition of the electrode prior to receipt.The device passed all of the electrical tests performed.However, the device had moisture content that exceeded the residual gas analysis test limit.Based on an assessment of the residual gas analysis data it is believed that this device was not hermetic.This older device configuration is no longer manufactured.This is the final report.
|