Event summary: the patient data files showed at least six injections were performed with catheter 2af284 / 12130-81 on the date of the event without any system notice.
Failure file did not show any system notice triggered on the date of the event.
Upon visual inspection of the flexcath sheath 4fc12 / 70121-055, results showed the device was intact with no apparent issues.
Air aspiration was reproduced when a lab test catheter was introduced through the sheath.
Dissection showed the hemostatic valve was leaking.
In conclusion, the reported air ingress issue was confirmed through testing but not confirmed through the data analysis.
The catheter passed the returned product inspection as per specification.
The sheath failed the returned product inspection due to a leaking hemostatic valve.
|
It was reported that during a cryo ablation procedure, the balloon catheter was inflated outside of the patient¿s body and air was a spirated.
When the balloon catheter was inserted into the sheath, air was introduced and ¿more resistance than usual¿ was observed.
It was noted that neither the balloon catheter nor the sheath were replaced.
The case was completed with cryo.
No patient complications have been reported as a result of this event.
|