MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAB ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAB ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAB
Device Problems Improper or Incorrect Procedure or Method (2017); Moisture or Humidity Problem (2986)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Currently it is unknown whether or not the device may have caused or contributed to the event. The device has been returned, but not yet evaluated. Further information will follow once the analysis has been completed. No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they fell in the lake. The insulin pump was beeping. There was fluid under the insulin pump¿s display. The insulin pump had no cracks. The customer¿s blood glucose level was unknown. The customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
Findings: pump was received with moisture damage on the lcd board, however unit passed the functional testing, including the operating currents measurement, self test, unexpected alarm error test, displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test. Pump had cracked case at display window corner, reservoir tube lip, minor scratched and cracked display window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name530G INSULIN PUMP MMT-751NAB
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6835995
MDR Text Key84627744
Report Number2032227-2017-43387
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/15/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/31/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMMT-751NAB
Device Catalogue NumberMMT-751NAB
Device Lot NumberA4751NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-