Model Number 407201
Device Problem Material Fragmentation (1261)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/27/2017
Event Type  malfunction  
Event Description
During the procedure, resistance was noted while inserting the needle in the sheath and a plastic shaving was noted.The device was used to complete the procedure and there were no adverse consequences to the patient.
 
Manufacturer Narrative
The results of the investigation concluded that the returned brk needle had been bent in multiple locations along its length and the needle curve no longer met specifications.The device met specifications prior to release from abbott manufacturing facilities as supported by a review of the device history record.The brk transseptal needle instructions for use (ifu) states, ¿do not alter this device in any way.¿ the cause of the reported complaint is consistent with altering the distal curve of the returned needle prior to insertion through the dilator/sheath assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TRANSSEPTAL NEEDLE
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6837361
MDR Text Key84306914
Report Number3008452825-2017-00190
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2020
Device Model Number407201
Device Catalogue Number407201
Device Lot Number5874946
Other Device ID Number05414734205160
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SWARTZ BRAIDED INTRODUCER
-
-