|
Model Number M001183020 |
Device Problems
Break (1069); Device Damaged Prior to Use (2284)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/10/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the catheter was broken.A renegade hi-flo kit was selected for use.During preparation, the physician flushed the lumen of the device with heparinized saline and made sure there were no leaks.However, when device was removed from the carrier hoop, it was noted that the catheter was broken into two parts.The procedure was completed with another of the same device.No patient complications reported and patient's status was stable.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.The hub, shaft and tip were microscopically examined.The device did not show a broken shaft; however, the shaft did show 2 kinks at the distal end of the device.The 1st kink was located approximately 5cm from the tip and the 2nd kink was located approximately 9cm from the tip.The alleged complaint of the broken shaft was not confirmed.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Event Description
|
It was reported that the catheter was broken.A renegade hi-flo kit was selected for use.During preparation, the physician flushed the lumen of the device with heparinized saline and made sure there were no leaks.However, when device was removed from the carrier hoop, it was noted that the catheter was broken into two parts.The procedure was completed with another of the same device.No patient complications reported and patient's status was stable.
|
|
Search Alerts/Recalls
|
|
|