• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA; SPINAL CORD STIMULATOR Back to Search Results
Model Number SC-8216-50
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Nerve Damage (1979)
Event Date 08/16/2017
Event Type  Injury  
Manufacturer Narrative
Additional suspect medical device component involved in the event: model #: sc-8216-70 serial #: (b)(4) description: artisan surgical lead, 70cm.
 
Event Description
A report was received that the patient was experiencing numbness in the right arm.X-ray result confirmed that the lead had migrated.The patient underwent a lead revision procedure wherein the lead was moved back into position and was sutured in.The patient was doing well postoperatively.
 
Manufacturer Narrative
Additional information was received that the physician did not suspect numbness was caused by the migrated lead but believed that positioning during surgery may have caused it.
 
Event Description
A report was received that the patient was experiencing numbness in the right arm.X-ray result confirmed that the lead had migrated.The patient underwent a lead revision procedure wherein the lead was moved back into position and was sutured in.The patient was doing well postoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key6841054
MDR Text Key84614597
Report Number3006630150-2017-03289
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729779902
UDI-Public08714729779902
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/23/2018
Device Model NumberSC-8216-50
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age41 YR
-
-