• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 4415
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/11/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Per the perfusionist, when the out valve was pushed in there was no water flow noting a possible scale and calcium obstruction.There were no fault lights indicated.The customer confirmed they descale the unit per protocol.The field service representative (fsr) was unable to duplicate the complaint.The arterial (art), cardioplegia (cpg) and blanket water lines were connected to the back of the heater cooler and the water flow was normal and strong.The customer was advised to descale the unit as sediment buildup could restrict the water flow.The unit operated to the manufacturer¿s specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure (cpb), the heater cooler's flow was lower than normal.It was noted to take an unusually long time when checking the cooling and warming modes.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6842924
MDR Text Key85453325
Report Number1828100-2017-00398
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 09/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received08/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/08/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-