• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM Back to Search Results
Model Number MT22495-PNK
Device Problems Use of Device Problem (1670); Invalid Sensing (2293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/05/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2017, that on (b)(6) 2017, there was an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter.The sensor was inserted at the abdomen on (b)(6) 2017.Reportedly, the patient did not calibrate after the inaccuracy.Additionally, bg values were not entered within 5 minutes of testing.No additional event or patient information is available.No product or data was provided for evaluation.The reported inaccuracy could not be confirmed.A root cause could not be determined.Labeling indicates: if the difference between your sensor glucose reading and blood glucose value is greater than 20% of the blood glucose value for sensor glucose readings > 80 mg/dl or greater than 20 points for sensor glucose readings < 80 mg/dl, wash your hands and take another blood glucose measurement.If the difference between this second blood glucose measurement and the sensor is still greater than 20% for sensor glucose readings > 80 mg/dl or greater than 20 points for sensor glucose readings < 80 mg/dl, recalibrate your sensor using the second blood glucose value.The sensor glucose reading will correct over the next 15 minutes.Labeling indicates: to calibrate the system, enter the exact blood glucose value that your blood glucose meter displays within 5 minutes of a carefully performed blood glucose measurement.Entering incorrect blood glucose values or blood glucose values from more than 5 minutes before entry might affect sensor performance, and you might miss a low or high blood glucose value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6843787
MDR Text Key85304440
Report Number3004753838-2017-77748
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10386270000160
UDI-Public10386270000160
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 08/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495-PNK
Device Catalogue NumberSTK-DR-PNK
Device Lot Number5226416
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/11/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
Patient Weight73
-
-