Siemens filed the initial mdr 1219913-2017-00192 on september 06, 2017.
09/08/2017 additional information: the type of type of blood collection tubes used by the customer are plastic gel barrier tubes manufactured in (b)(4).
The customer serially diluted (x3 manual dilution) only sample 1.
The result was 255 ng/ml.
The patient sample cannot be returned to the manufacturer's site due to china customs issue.
Therefore, no further investigation is possible.
The patient was healthy, no cancer, and it is unknown if the patient is taking any medications.
The cause for the discordant afp results is unknown.
The instrument is performing within specifications.
No further evaluation of the device is required.
Mdr 1219913-2017-00201 supplemental report 1 was filed for the same event.
|
The cause for the discordant afp results is unknown.
Siemens healthcare diagnostics is investigating.
The ifu states in the warning section: "warning the concentration of afp in a given specimen, as determined by assays from different manufacturers, can vary due to differences in assay methods and reagent specificity.
The results reported by the laboratory to the physician must include the identity of the afp assay used.
Values obtained with different afp assay methods cannot be used interchangeably.
Before changing assay methods, the laboratory must do the following: -for prenatal testing, the laboratory must establish a range of normal values for the new assay based on normal serum and amniotic fluid from pregnant women with a confirmed gestational age.
-for cancer management, the laboratory must perform additional testing to confirm baseline values for patients being serially monitored.
United states federal law restricts this device to sale and distribution by or on the order of a physician, or to a clinical laboratory; and use is restricted to, by, or on the order of a physician.
Use afp results only as part of the overall clinical evaluation of a patient.
Do not use afp results as the only criterion for diagnosis.
" the ifu states in the interpretation of results section: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings.
" the ifu states in the limitations section: "heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays.
Patients routinely exposed to animals or to animal serum products can be prone to this interference and anomalous values may be observed.
Additional information may be required for diagnosis.
The advia centaur afp assay uses antibodies from two animal species and routinely adds animal sera to its assay components to minimize the interfering response.
Mdr 1219913-2017-00201 was filed for the same event.
|