Report received that high impedance was found after performing system diagnostics on new generator in the operation room.The patient was scheduled for a generator replacement due to battery depletion, and the generator was unable to be interrogated.When the new generator was implanted and connected to the lead, a system diagnostic was run and showed high impedance.The lead was explanted.Further information was received that during the surgery, the lead pin was confirmed to be inserted into the generator properly.Generator diagnostics were also performed on the new generator and the results showed normal impedance.A review of the programming history did not indicate whether or not high impedance was seen prior to the replacement.Recent system diagnostics prior to surgery could not be provided by the patient's physician.The lead and generator were explanted and received by the manufacturer, but analysis has not been approved to date.No other relevant information has been provided to date.
|
Further information was received that product analysis was completed on the lead.The lead was returned in two portions.Visual examination showed one portion had stretched and kinked quadfilar coils.On this portion, set screw marks were present and white deposits were observed in various locations.Abrasions were also noted on the connector boot.Dried remnants of what appeared to be bodily fluids were inside the outer silicone tubing.Other abraded openings were seen along the outer tubing of this portion.On the second portion, white deposit was also seen.Besides typical markings and damage associated with explant, no visual anomalies were noted.No discontinuities were found during the continuity checks for the returned portions.However, there was a portion of the lead which included the electrodes which was not returned for analysis and an assessment on that portion could not be made.No further relevant information has been received to date,.
|