Findings: no blue display or full missing segment noted.Unit received with blank display due to moisture damaged electronic assembly.Also moisture damage motor and vibrator during visual inspection.Unable to perform operating current test, unexpected alarm error test or all functional testing including the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test due to blank display.Unit had minor scratched lcd window, cracked lcd window, cracked case at display window corners, cracked battery tube threads and cracked reservoir tube lip.
|