• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAS
Device Problems Moisture Damage (1405); Improper or Incorrect Procedure or Method (2017); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump was exposed to moisture.The customer¿s blood glucose was unknown at the time of the incident.The customer states the insulin pump was exposed to fluid when he fell into a pool.Customer states that he put in a new battery and the blue came on.Nothing else came up for about five minutes and then gave an alarm.Customer then pulled battery out.Customer reports there is fluid under the lcd display.Customer states pump was without battery for two weeks to allow it to dry.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Findings: no blue display or full missing segment noted.Unit received with blank display due to moisture damaged electronic assembly.Also moisture damage motor and vibrator during visual inspection.Unable to perform operating current test, unexpected alarm error test or all functional testing including the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test due to blank display.Unit had minor scratched lcd window, cracked lcd window, cracked case at display window corners, cracked battery tube threads and cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6850602
MDR Text Key85298890
Report Number2032227-2017-45762
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503717
UDI-Public(01)00643169503717
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAS
Device Catalogue NumberMMT-751NAS
Device Lot NumberA4751NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received11/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age59 YR
Patient Weight238
-
-