• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE,MOT DR UNIT, HNDCONT POWERMAX ELITE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE,MOT DR UNIT, HNDCONT POWERMAX ELITE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 72200872S
Device Problems Overheating of Device (1437); Failure to Shut Off (2939)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation. There was a relationship found between the returned device and the reported incident. Corrosion was observed on the cable assembly connection and gearbox fork assembly. Product failed functional testing with motor stall error and overheating. Cause of overheating and errors is a corroded motor/gearbox. The motor/gearbox assembly could not be removed from the housing for further assessment due to corrosion. The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly. A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing. Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended. A corrective action has been initiated to mitigate future recurrence of similar events.
 
Event Description
A report of extremely hot has been received. The customer states when plugged in the shaver wouldn't turn off. No injury was reported. This did not take place during a procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVCE,MOT DR UNIT, HNDCONT POWERMAX ELITE
Type of DeviceINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 w. william cannon drive
austin, TX 78735
MDR Report Key6853233
MDR Text Key135538939
Report Number1643264-2017-00528
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/19/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200872S
Device Lot NumberAAO1684
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-