BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿ CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
|
Back to Search Results |
|
Model Number H7493808012450 |
Device Problems
Kinked (1339); Device Damaged Prior to Use (2284)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/29/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
Device evaluated by mfr.
: returned product consisted of a quantum maverick balloon catheter.
The balloon was loosely folded.
The outer shaft, inner shaft, balloon and tip were microscopically examined.
The tip is damaged.
The hypotube shaft was completely separated 75.
8cm from the hub.
The fracture faces were oval as if kinked prior to separation.
There was no evidence of any material or manufacturing deficiencies contributing to the damage.
There were numerous hypotube and shaft kinks.
There is no evidence that the device failed to meet applicable product specifications prior to shipment.
The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.
(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2017.
It was reported that shaft kink occurred.
A 4.
5mm x 12mm quantum maverick balloon catheter was selected for use.
However, the delivery shaft was noted to be kinked during unpacking.
The procedure was completed with another of the same device.
No patient complications were reported and patient's status was stable.
However, returned device analysis revealed hypotube break.
|
|