CINCINNATI SUB-ZERO, LLC. HEMOTHERM 400 CE; CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER. Back to Search Results
Model Number 400CE
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/15/2017
Event Type  malfunction  
Manufacturer Narrative
Csz medical technical support received a complaint stating the hemotherm shut down during a procedure.There was no patient harmed or injured and the procedure continued successfully.The devices power supply board was returned to csz for investigation.Csz found discoloration and the board to be inoperative.The power supply board was replaced and the device functions as designed.
 
Event Description
Cincinnati sub-zero medical technical support received a complaint stating the hemotherm shut down during a procedure.There was no patient or user injury reported and the procedure was able to continue successfully.This report was filed in our complaint handling system as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOTHERM 400 CE
Type of Device
CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER.
Manufacturer (Section D)
CINCINNATI SUB-ZERO, LLC.
12011 mosteller road
cincinnati OH 45241
Manufacturer Contact
christina miracle
12011 mosteller road
cincinnati, OH 45241
MDR Report Key6857885
MDR Text Key85783883
Report Number1516825-2017-00017
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122813
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 09/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number400CE
Device Catalogue Number86022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-