It was reported that during a da vinci assisted total benign hysterectomy procedure, there was vision loss in the right eye.
The intuitive surgical, inc.
(isi) technical support engineer (tse) advised the customer to ensure the cable connections were tight and perform a system restart with breaker reset; however, the issue persisted.
The tse reviewed the logs and found an error 48232 pointing to the dual camera controller unit (doco).
The tse had the customer remove the camera cable from the doco, power system up and install a new camera cable.
The error 48232 recurred.
The surgeon made the decision to convert and complete the procedure as a traditional open surgical procedure.
There was no patient harm, adverse outcome or injury reported.
An isi field service engineer (fse) was dispatched to the facility and replaced the doco to resolve the issue.
The doco is located on the vsc and it receives and processes video signals.
|