SYNTHES OBERDORF COLIBRI II INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.101 |
Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/23/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.
During repair, it was determined that the control unit was not functioning and defective.
Therefore, the reported condition was confirmed.
The assignable root cause was determined to be due to improper maintenance which is user error/misuse/abuse.
If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
(b)(4).
(b)(6).
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the small battery drive device was working on and off; and that it had a bad contact.
There were no delays in the surgical procedure.
It was not reported whether a spare device was available for use.
There was patient involvement.
There were no injuries, medical intervention or prolonged hospitalization.
All available information has been disclosed.
If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|