The investigation determined that higher than expected vitros na+ results were obtained from a single non-vitros biorad l1 quality control fluid, lot 31820, using vitros na+ slides on a vitros 4600 chemistry system.The event was isolated to the calibration event performed on 19 august 2017.In addition to the vitros na+ assay, the vitros k+ and cl- assays were also affected.The calibrations were suboptimal when compared to the typical calibration responses and parameters.The most likely cause of the suboptimal 19 august calibrations was user error, where the reconstituted calibrators (calibrator kit 257) were stored for too long at room temperature prior to use.Acceptable vitros na+ performance was observed after a recalibration event was performed using properly handled calibrator fluids.The investigation found no indication the vitros 4600 chemistry system or vitros na+ slide lot contributed to the event.
|
The investigation determined that higher than expected vitros na+ results were obtained from a single non-vitros biorad l1 quality control fluid, lot 31820, using vitros na+ slides on a vitros 4600 chemistry system.Biorad l1 na+ results of 138.9, and 142.7 mmol/l vs.The expected result of 126.0 mmol/l.Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples.A review of vitros na+ results during the timeframe of the event determined that two patient samples were affected and higher than expected results were reported outside of the laboratory.However, corrected reports were issued and there were no allegations of patient harm as a result of this event.This report corresponds to ortho clinical diagnostics inc.Complaint number (b)(4).
|