Catalog Number C-HS-3045 |
Device Problem
Failure to Align (2522)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/30/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).A lot history record review was completed for the reported product lot number.There was no nonconformance recorded in the lot history.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.(b)(4).A lot history record review was completed for the reported product lot number.There was no nonconformance recorded in the lot history.The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
|
|
Event Description
|
The hospital reported that during preparation for a coronary artery bypass procedure, hs iii proximal seal was not loaded into the delivery system properly.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
|
|
Manufacturer Narrative
|
(b)(4).The device was returned to the factory for evaluation.A visual inspection was conducted.Signs of clinical use and evidence of blood were observed on both the loading device and the delivery device.The delivery device was returned outside the loading device.The white plunger was depressed on the delivery device.The blue slide lock was disengaged.The tension spring assembly remained inside the delivery tube with the seal extended outside the delivery tube completely unfolded.The seal and tension spring were removed from the delivery tube for inspection.Microscopic inspection determined that the seal, though tainted with blood, was intact, with no cracks or delamination.Blood was visible in the delivery device indicating that there was attempt to deploy the device into the aorta.Based on the received condition of the device, we were not able to measure the delivery tube dimensions.Based on the returned condition of the device, it was apparent that the seal was properly loaded.¿ the reported failure mode "fitting problem" was therefore not confirmed.
|
|
Event Description
|
The hospital reported that during preparation for a coronary artery bypass procedure, hs iii proximal seal was not loaded into the delivery system properly.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
|
|
Search Alerts/Recalls
|