• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) UPGRADE PRGM 9734261 STEALTHSTATION S7; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) UPGRADE PRGM 9734261 STEALTHSTATION S7; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Patient Involvement (2645)
Event Date 08/25/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.No parts were replaced.No parts have been received by manufacturer for evaluation.Part not returned for evaluation.
 
Event Description
A medtronic representative reported that while outside of surgery the surgeon monitor of the navigation system was flickering.Checking the limo connector did not resolve the issue.There was no patient present at the time of the issue.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The representative was unable to replicate the reported issue.The representative reported that all connections were checked and the navigation system was powered on for 45 minutes without issue.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPGRADE PRGM 9734261 STEALTHSTATION S7
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key6886219
MDR Text Key87877552
Report Number1723170-2017-03871
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 10/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9734261
Is the Reporter a Health Professional? No
Date Manufacturer Received09/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-