A system diagnostic test revealed high impedance on a patient's device.The following day, the patient underwent lead and generator replacement surgery, and preoperative diagnostic testing also showed high impedance.During the surgery, the physician confirmed that the lead pin had been fully inserted into the generator, leading him to believe that a lead fracture was the cause of the high impedance.The explanted lead was returned to the manufacturer for analysis, but analysis has not been approved to date.No additional relevant information has been received to date.
|
Analysis was approved for the lead.The lead was returned in two portions; however, the lead electrodes were not returned for evaluation so an assessment could not be made on that portion of the lead.Two pairs of set screw marks were observed on the connector pin, providing evidence that proper contact between the set screws twice.A single set screw indentation was observed at the end tip of the connector pin suggesting that the lead connector pin was not inserted completely at one point.An abrasion was also observed on the connector boot, and several abraded openings were observed on the outer and inner tubing in multiple locations.A coil break was also identified, and microscopy suggested that a stress-induced fracture occurred in the negative coil.No additional relevant information has been received to date.
|