PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; JCX Back to Search Results
Catalog Number PMX110
Device Problems Pumping Stopped (1503); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/31/2017
Event Type  malfunction  
Manufacturer Narrative
Results: a pipe cleaner was inserted into the penumbra system aspiration pump max 110 (pump max) vacuum port, and blood was observed to be inside the pump.Conclusion: evaluation of the returned device revealed that the pump max had blood inside.This type of damage typically occurs due to improper handling during use.If the aspiration tubing is connected directly to the vacuum inlet rather than the canister supplied by penumbra, blood will likely enter the pump assembly.The blood inside the pump max likely caused the pump max to stop working during the procedure.Penumbra pumps are visually inspected and 100% functionally tested during incoming inspection by quality.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a thrombectomy procedure using penumbra system aspiration pump max 110 (pump max).During the procedure, the pump max stopped working.Although there was no report of an adverse effect on the patient, multiple attempts have been made to gather additional information.As of 25-sept-2017, no additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key6892212
MDR Text Key89257130
Report Number3005168196-2017-01646
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 08/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPMX110
Device Lot NumberF07836-07
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2017
Date Manufacturer Received08/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-