Catalog Number 72203013
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/04/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the inner blade broke during the procedure.No pieces were introduced into the patient.A backup device was available to complete the procedure with minimal delay.No patient impact was reported.
 
Manufacturer Narrative
One 72203013 incisor plus platinum 4.5 mm blade returned.This device is in damaged condition.There is a bow of the overall length of the outer blade.There are wear bands on the outer surface of the inner blade.The sluff chamber of the inner blade has attained pressure cracks.These conditions are symptoms of force and excessive side load applied.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM INCISOR PLUS PLATINUM BLADE
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key6897226
MDR Text Key87727478
Report Number1219602-2017-01213
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72203013
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
-
-