SMITH & NEPHEW, INC. SVC, CAMERA HEAD, NON-AC, HD560H; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 72200561S
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2017
Event Type  malfunction  
Event Description
It was reported the camera head non-ac hd560h picture blacks out when used.A back up device was utilized to complete the procedure.No patient injury or complications were reported.
 
Manufacturer Narrative
There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of loss of live video could not be reproduced.Product passed functional testing during 6 hour burn-in with no signal loss or black outs.Video image remained constant throughout burn-in.All functions perform as expected.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the functional testing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC, CAMERA HEAD, NON-AC, HD560H
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 w. william cannon dr.
austin, TX 
MDR Report Key6903684
MDR Text Key87773558
Report Number1643264-2017-00589
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier03596010606808
UDI-Public(01)03596010606808
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070266
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number72200561S
Device Catalogue Number72200561S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received10/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-