AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B1120-020
Device Problems Detachment Of Device Component (1104); Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/12/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4) it is unknown if the device is returning.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat a lesion that was previously stented located in the right common iliac vein.The armada 35 12 mm / 20 mm balloon catheter ruptured during the first inflation at 10 atmospheres.A piece of the balloon separated where the marker was but stuck to the distal end of the guide wire and was easily removed through the sheath.The instructions for use were followed with no issues noted during preparation.There was no resistance during advancement or withdrawal of the device.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was not returned for evaluation.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported difficulties.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6909268
MDR Text Key88115452
Report Number2024168-2017-07920
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Catalogue NumberB1120-020
Device Lot Number50408G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
Patient Weight106
-
-