MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. MODULAR HANDLE, T RATCHETING TLR 90 IN-LBS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. MODULAR HANDLE, T RATCHETING TLR 90 IN-LBS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 3356-1
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 09/06/2017
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event Description
It was reported that the torque output of a torque limiting handle was found outside of the adequate performance range.This was detected during a preliminary calibration check by zimmer biomet spine personnel.There are no specific surgical procedures associated with this handle.
 
Manufacturer Narrative
Additional information: the returned handle was evaluated.The torque output was confirmed to be below the acceptable limit.The cause cannot be definitively determined, but typical contributors include spring relaxation, wear to critical components, and a breakdown of internal lubrication over multiple uses.A review of the manufacturing records did not identify any issues which would have contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR HANDLE, T RATCHETING TLR 90 IN-LBS
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key6911937
MDR Text Key88127071
Report Number3012447612-2017-00536
Device Sequence Number1
Product Code HXC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number3356-1
Device Lot Number86KC-079
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-