• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION AG-920RA; MULTI-GAS UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION AG-920RA; MULTI-GAS UNIT Back to Search Results
Model Number AG-920RA
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/07/2017
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer reported that when the doctor attempted to use the multi gas unit, the monitor wasn't reading the oxygen value correctly.The biomedical engineer sent in the unit for evaluation and requested a loaner.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The biomedical engineer reported that when the doctor attempted to use the multi gas unit, the monitor wasn't reading the oxygen value correctly.
 
Manufacturer Narrative
The biomedical engineer reported that when the doctor attempted to use the multi gas unit, the monitor wasn't reading the oxygen value correctly.The biomedical engineer sent in the unit for evaluation and requested a loaner.The device was cleaned and evaluated.The reported problem of "inaccurate reading" was duplicated.All malfunctioning parts were replaced.The unit was tested per operator's/service manual and completed 24 hours of extended testing.The unit operates to manufacturer's specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AG-920RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key6914157
MDR Text Key88163361
Report Number8030229-2017-00328
Device Sequence Number1
Product Code CCK
UDI-Device Identifier00841983103162
UDI-Public00841983103162
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/29/2017,09/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAG-920RA
Device Catalogue NumberAG-920RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Report Sent to FDA09/29/2017
Distributor Facility Aware Date05/24/2016
Device Age91 MO
Event Location Hospital
Date Report to Manufacturer09/29/2017
Date Manufacturer Received09/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-