Model Number 16-02-85
Device Problems Failure to Power Up (1476); Electrical Power Problem (2925)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/11/2017
Event Type  Malfunction  
Manufacturer Narrative

Patient information was not provided. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). This medwatch report is being filed on behalf of livanova (b)(4). A livanova field service representative was dispatched to the facility to investigate. The service representative was unable to confirm the reported failure. The compressor was tested and was found to function as expected and the system compressor turned on when required. The facility biomedical engineer indicated that the relay was the issue, which the customer had already replaced. A performance test and technical safety inspection were completed without issue and the unit was returned to service. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. As the issue could not be reproduced, a root cause was not determined. However, based on the report from the biomedical engineer, the likely root cause was a faulty relay. Issue resolved by customer.

 
Event Description

Livanova (b)(4) received a report that the compressor of the heater-cooler system 3t did not activate when the system was turned on. The device tripped the circuit which shut the system off. This issue was noted post-procedure. There was no report of patient injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6916083
MDR Text Key88467015
Report Number9611109-2017-00775
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation HEALTH PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 10/04/2017
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/05/2017
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-85
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/11/2017
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured06/25/2014
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-