• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems Connection Problem (2900); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/22/2017
Event Type  malfunction  
Manufacturer Narrative
Product event summary: manufacturer's analysis confirmed the customer comment that there was a problem with the connector of the device header; the connection was loose on the link electronic module (lem) board.It was also noted that the front keyboard latch was broken, cord bay was broken, errors were found in the software history, and the emergency button touch screen did not work.All found defective parts were replaced and all other identified issues were resolved.The device passed final functional and system tests.
 
Event Description
It was reported that there was a problem with the radio frequency (rf) programmer head connector on the programmer.The programmer has been returned for repair.No patient complications have been reported as a result of this event.It was further reported that the programmer originally returned as an exchange subsequently tested out of specification during manufacturer¿s analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6920869
MDR Text Key89507235
Report Number2182208-2017-01541
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received07/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-