• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED XMTR MMT-7811NA GST3C LOCKOUT; SENSOR, GLUCOSE, INVASIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED XMTR MMT-7811NA GST3C LOCKOUT; SENSOR, GLUCOSE, INVASIVE Back to Search Results
Model Number MMT-7811NA
Device Problems Failure to Charge (1085); Moisture Damage (1405); Battery Problem (2885); Moisture or Humidity Problem (2986)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/17/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump had moisture.The customer¿s blood glucose level was unknown.Customer stated that transmitter will no longer charge and not sure of how long moisture exposure was for.The customer was assisted with troubleshoot and customer reports that the transmitter might have become damaged from the hurricane and may have been submerged in water.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device received with moisture damage at connector pins.Unable to charge or perform functional test due to contamination at connector pins.
 
Manufacturer Narrative
The event was reported in error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XMTR MMT-7811NA GST3C LOCKOUT
Type of Device
SENSOR, GLUCOSE, INVASIVE
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire st.
northridge, CA 91325-1219
MDR Report Key6921190
MDR Text Key89398759
Report Number2032227-2017-56409
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup,Followup
Report Date 04/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-7811NA
Device Catalogue NumberMMT-7811NA
Device Lot NumberA000390531
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received04/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age33 YR
Patient Weight227
-
-