It was reported that during the implant procedure, right ventricular (rv) lead oversensing was noted after the cardiac resynchronization therapy pacemaker (crt-p) was connected.
The rv oversensing could be replicated with header manipulation when the device was in the pocket, but could not be replicated when the device was outside the pocket.
The rv lead was removed from the device and reconnected, but similar rv oversensing was still noted.
The rv lead was disconnected and tested with an analyzer and electrical parameters were the same as those prior to the original connection of the device.
The physician manipulated the proximal end of the lead, including tie-down sleeve, and no oversensing was noted.
It was concluded that the problem was in the header of the can and not the rv lead itself.
The device was not used and replaced with no oversensing noted.
No patient complications have been reported as a result of this event.
|