Model Number U128 |
Device Problems
Premature Discharge of Battery (1057); High impedance (1291); Connection Problem (2900)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/06/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time, the device has not been returned.If the lead is returned, analysis will be performed and this report would be updated at that time.
|
|
Event Description
|
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p) exhibited right ventricular (rv) impedance measurements of less than 200 ohms and greater than 3000 ohms.The threshold measurements were also increased.The device also appeared to be prematurely depleting, and the physician thought there was a possible set screw issue with the rv lead.A surgical revision was performed and the patient's rv lead was tested via the pacing system analyzer (psa).The high impedances were noted with the psa as well.The device was explanted and replaced.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the product was performed.Review of device memory confirmed the presence of a higher than normal current drain condition.Electrical testing and analysis were then conducted, which isolated the high current to an anomaly in an oxide layer within a custom integrated circuit component.This anomaly caused a high current drain, which over time resulted in the reported clinical observations.
|
|