GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problems Premature Discharge of Battery (1057); High impedance (1291); Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/06/2017
Event Type  Injury  
Manufacturer Narrative
At this time, the device has not been returned.If the lead is returned, analysis will be performed and this report would be updated at that time.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p) exhibited right ventricular (rv) impedance measurements of less than 200 ohms and greater than 3000 ohms.The threshold measurements were also increased.The device also appeared to be prematurely depleting, and the physician thought there was a possible set screw issue with the rv lead.A surgical revision was performed and the patient's rv lead was tested via the pacing system analyzer (psa).The high impedances were noted with the psa as well.The device was explanted and replaced.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the product was performed.Review of device memory confirmed the presence of a higher than normal current drain condition.Electrical testing and analysis were then conducted, which isolated the high current to an anomaly in an oxide layer within a custom integrated circuit component.This anomaly caused a high current drain, which over time resulted in the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key6932233
MDR Text Key88841910
Report Number2124215-2017-14824
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/22/2019
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received09/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age45 YR
-
-