• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Output below Specifications (3004)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/18/2017
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date.The stm replaced the scroll compressor as per the recommendations in the service manual to fix the reported issue.A full system test and preventative maintenance was complete, and all functionality and safety tests were performed and passed according to factory specifications.The iabp was returned to the customer and cleared for clinical use.(b)(6).
 
Event Description
It was reported that during scheduled preventive maintenance (pm) of the intra-aortic balloon pump (iabp) by the getinge service territory manager (stm), it was found that the compressor could not meet the required test criteria (pressure below specification).The iabp's fault logs also recorded multiple error code #77 which indicates compressor replacement was required per the service manual there was no patient involved and no death, injury nor adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key6932787
MDR Text Key90157608
Report Number2249723-2017-00611
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Device AgeYR
Date Manufacturer Received09/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-