GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problems High impedance (1291); Ambient Noise Problem (2877)
Patient Problem Syncope (1610)
Event Date 09/28/2017
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p), and another manufacturer's right ventricular (rv) lead, exhibited high out of range pacing impedance measurements and noise.It was noted the patient was undergoing a procedure at the same time the high impedance measurements were recorded.It was reported that the patient is pacer dependent and had experienced some syncope due to noise on the rv lead.The device had switched to unipolar due to the out of range measurements.The device was reprogrammed back to bipolar and measurements were acceptable.No further changes are planned at this time.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key6936053
MDR Text Key89107001
Report Number2124215-2017-18192
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 09/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/27/2017
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received09/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age74 YR
-
-