SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616SR
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/27/2017
Event Type  malfunction  
Manufacturer Narrative
There was no relationship found between the returned device and the reported incident. A visual inspection was performed on the product and no issue was observed. Product passed functional and high speed testing (100-10,000 rpms) for 5 minutes in oscillate, forward, and reverse directions with no sticking buttons. Unit passed functional tests on the dyonics power, dii and dii eip test control units with and without footswitch. Current draw was average for this product and all 3 buttons performed as expected during functional testing. The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the functional testing process. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Event Description
It was reported the powermax elite mdu hand control buttons were sticking. No patient injury or complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVCE REPL MDU CNTRL PWRMX ELITE
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 w. william cannon dr.
austin, TX 78735
MDR Report Key6939793
MDR Text Key139548903
Report Number1643264-2017-00616
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/23/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200616SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-