US SURGICAL PUERTO RICO PROTACK LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 174006
Device Problems Failure to Advance (2524); Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/25/2017
Event Type  Malfunction  
Event Description

According to the reporter, during a ventral hernia repair procedure, after firing some clips jammed. The edge of the clip remained outside of the shaft. There was no patient injury.

 
Manufacturer Narrative

Evaluation summary post market vigilance (pmv) led an evaluation of one device. Visual inspection of the instrument noted that the handle was detached from the rod. A tack was protruding from the shaft. Functional testing was precluded due to the observed condition of the instrument. A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture. Replication of the observed conditions may occur if the instrument is handled roughly during use. The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident. Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate. If information is provided in the future, a supplemental report will be issued.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROTACK
Type of DeviceLAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 midletown ave
north haven, CT 06473
2034925563
MDR Report Key6947813
MDR Text Key89304888
Report Number2647580-2017-07168
Device Sequence Number1
Product Code GCJ
Combination Product (Y/N)N
Reporter Country CodeGR
PMA/PMN NumberK090470
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type FOREIGN,OTHER,USER FACILITY
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/05/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/13/2017
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number174006
Device Catalogue Number174006
Device LOT NumberP6A0617X
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer10/20/2017
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/15/2018
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured01/25/2016
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-