MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 09/18/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. A medtronic representative went to site to test the equipment. Representative reported that a monitor was changed and the issue was resolved. A system checkout was performed and the hardware, software, and instruments passed the system checkout. The system was found to be fully functional. The suspect monitor has been received by manufacturer for evaluation. The received monitor had no led when connected to power source. Monitor had scratches on upper and lower left of screen. The hardware investigation found that the reported event was related to a hardware issue. This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Event Description
A medtronic representative reported that during outside of procedure, while inputting network configurations after installing the system, the camera cart screen of the navigation system displayed a purple distorted screen and the monitor turned off. Representative plugged the main monitor cables to camera cart screen but the issue was not resolved. But when the camera cart monitor cables were plugged into the main cart monitor, the display was functioning normally. Self test did not indicate any failures. There was no patient present at the time of the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S8 PREMIUM SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key6949741
MDR Text Key90088936
Report Number1723170-2017-04149
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 10/13/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number9735665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-